Introduction:
|
I-MAX TOUCH CEPH
|
Cephalometry: Module dedicated to orthodontic exams
The I-Max Touch produces cephalometric images of irreproachable quality (High-Definition), and with a significant reduction in the X-ray dose. With the cephalometric option, benefit from 17 programs in total to acquire any diagnosis x-ray, needed for the art of dentistry:
|
- Standard adult or child panoramic
- Left or right semi-panoramic
- Reduced dose panoramic
- Bitewing mode
- Incisive block
- Maxillary sinus
- Panoramic with improved orthogonality
- TMJ open/closed mouth
- 4 quadrants / Tomographic slices
Ceph Programs ceph: A choice of 5 formats
- Ceph frontal : 24 x 22 cm
- Ceph lateral : 18 x 22 cm, 24 x 22 cm, 30 x 22 cm
- Carpus
|
|
Mobile sensor: cephalometry in flexible and affordable mode
The choice, depending on the use (panoramic or cephalometric), of a single sensor mode or with two sensors. Each sensor can be moved using its ergonomic handle that facilitates the transfer.
TECHNICAL SPECIFICATIONS
I-MAX TOUCH (Ceph Option)
- HF Generator At constant potential
- Focal spot 0.5 EN60336
- Min. total filtration 2.5 mm Al equivalent
- Anode voltage 60 - 86 kV
- Anode current 6 - 10 mA PAN, 6 - 12 mA CEPH
- Exposure time Pan 13.8 s / Ceph starting at 4.5 s
- Column Motorized telescopic
- SID (Source to Image Distance) Pan 500 mm / Ceph 1650 mm
- Digital sensor HD CCD + optical fiber plate
- CCD resolution 10.4 lp/mm
- Connection Direct acquisition (network cable) and/or integrated touch screen
- Storage Computer and/or USB memory stick
- Power supply 100-120 V, 220-240 V, 50/60Hz
- Amperage 8 A
Sensor data
- Digital sensor Amorphous silicon flat panel
- Surface 130 x 130 mm, 512 x 512 pixels
- F.O.V. (Filed Of View) 43 x 93 mm / 53 x 93 mm / 43 x 93 mm
- 3D volume voxel size 92 μm
|
|